Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
85
Uploading Data
1. Double click on SeatermV2.exe. The main screen appears.
2. In the Instruments menu, select SBE 19plus V2. Seaterm232 opens.
3. Seaterm232 will try to automatically connect to the 19plus V2. As it
connects, it sends GetHD and displays the response. Seaterm232 also fills
the Send Commands window with the correct list of commands for your
19plus V2. If there is no communication:
A. In the Communications menu, select Configure. The Serial Port
Configuration dialog box appears. Select the Comm port and baud
rate for communication, and click OK.
B. In the Communications menu, select Connect (if Connect is grayed
out, select Disconnect and reconnect).
C. If there is still no communication, check cabling between the
computer and 19plus V2.
D. If there is still no communication, repeat Step A with a different baud
rate and/or comm port, and try to connect again. Note that the
factory-set baud rate is documented on the Configuration Sheet.
4. Command the 19plus V2 to stop data logging by the method applicable to
your instrument’s setup:
• Pressing the Enter key, typing Stop, and pressing the Enter key again,
or
• Moving the magnetic switch to the Off position (only applicable in
Profiling mode, if IgnoreSwitch=N)
5. Display 19plus V2 status information by typing DS and pressing the Enter
key. The display looks like this (if in Profiling mode):
SBE 19plus V 2.3 SERIAL NO. 4000 15 Mar 2011 14:02:13
vbatt = 10.1, vlith = 8.9, ioper = 61.9 ma, ipump = 20.8 ma,
iext01 = 76.2 ma
status = not logging
number of scans to average = 1
samples = 5000, free = 4381542, casts = 1
mode = profile, minimum cond freq = 3000, pump delay = 60 sec
autorun = no, ignore magnetic switch = no
battery type = alkaline, battery cutoff = 7.5 volts
pressure sensor = strain gauge, range = 1000.0
SBE 38 = no, WETLABS = no, OPTODE = no, Gas Tension Device = no
Ext Volt 0 = yes, Ext Volt 1 = yes
Ext Volt 2 = no, Ext Volt 3 = no
Ext Volt 4 = no, Ext Volt 5 = no
echo characters = yes
output format = converted decimal
output salinity = no, output sound velocity = no
Verify that the status shows status = not logging.
Data may be uploaded during
deployment or after recovery. If
uploading after recovery, connect the
I/O cable as described in Section 3:
Power and Communications Test.
You may need to send Stop several
times to get the 19plus V2 to respond.