Manual revision 006 Section 1: Introduction SBE 19plus V2
5
Section 1: Introduction
This section includes a Quick Start procedure, and photos of a standard
SBE 19plus V2 shipment.
About this Manual
This manual is to be used with the SBE 19plus V2 SEACAT Profiler
Conductivity, Temperature, and Pressure Recorder. It is organized to guide the
user from installation through operation and data collection. We have included
detailed specifications, command descriptions, maintenance and calibration
information, and helpful notes throughout the manual.
Sea-Bird welcomes suggestions for new features and enhancements of our
products and/or documentation. Please contact us with any comments or
suggestions ([email protected] or 425-643-9866). Our business hours are
Monday through Friday, 0800 to 1700 Pacific Standard Time (1600 to 0100
Universal Time) in winter and 0800 to 1700 Pacific Daylight Time (1500 to
0000 Universal Time) the rest of the year.
Quick Start
Follow these steps to get a Quick Start using the SBE 19plus V2.
The manual provides step-by-step details for performing each task:
1. Install batteries and test power and communications (see Section 3: Power
and Communications Test).
2. Deploy the 19plus V2 (see Section 4: Deploying and Operating
SBE 19plus V2):
A. Install new batteries if necessary.
B. Ensure all data has been uploaded, and then send InitLogging to
make entire memory available for recording if desired.
C. Set date and time (DateTime=) and establish setup and logging
parameters.
D. Check status (DS) and calibration coefficients (DCal) to verify setup.
E. Moored mode (MM) - Set 19plus V2 to start logging now
(StartNow) or in the future (StartDateTime= and StartLater).
F. Install dummy plugs and/or cable connectors, and locking sleeves.
G. If applicable, remove Tygon tubing that was looped end-to-end
around conductivity cell for storage. Reconnect tubing from pump to
conductivity cell.
H. Profiling mode (MP) - As appropriate for your instrument’s setup:
- Put magnetic switch in On position, OR
- (if IgnoreSwitch=Y) Send commands to start logging now
(StartNow) or in the future (StartDateTime= and StartLater), OR
- (if AutoRun=Y) Apply external power.
I. Deploy 19plus V2.