Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2
22
Shown below is the vertical mount plumbing arrangement of a 19plus V2
configured with the optional SBE 5T titanium or 5P plastic pump, SBE 43
dissolved oxygen (DO) sensor, and SBE 18 pH sensor. Note that the SBE 43 is
plumbed into the system between the conductivity cell outlet and the Y-fitting.
The SBE 18 is not connected to the plumbing.
Note:
Sea-Bird recommends use of
the pH sensor on the 19plus V2
only when the 19plus V2 is in
Profiling mode (MP).
Air bleed valve detail
Note: Periodically clean air
bleed valve to ensure
proper functioning.
0.5 mm (0.02 in.)
Main plumbing is 13 mm ID x 19 mm OD
(1/2 inch x 3/4 inch) Tygon tubing.
A 13 mm (1/2 inch) long piece of
9.5 mm ID x 13 mm OD (3/8 inch x 1/2 inch)
Tygon tubing to provide tight seals to the main
plumbing is installed:
- on the SBE 43 intake and exhaust, and
-
on the conductivity cell exhaust when optional
anti-foul fittings are installed on the 19plus V2.