Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
41
Status Commands (continued)
GetSD Get and display status data, which contains
data that changes while deployed.
List below includes, where applicable,
command used to modify parameter:
Device type, Serial number
Date and time [DateTime=] in
ISO8601-2000 extended format
(yyyy mm-ddThh:mm:ss)
Logging status (not logging, logging,
waiting to start at . . ., or unknown status)
Number of recorded events in event
counter [reset with ResetEC]
Voltages and currents -
- Main battery voltage
- Back-up lithium battery voltage
- Operating current
- Pump current
- External voltage sensor current (channels
0 and 1) displays only if 1 or more
channels enabled
- External voltage sensor current (channels
2, 3, 4, and 5) displays only if 1 or more
channels enabled
- RS-232 sensor current displays only if
channel enabled
Memory - [reset with InitLogging]
- Number of bytes in memory
- Number of samples in memory
- Number of additional samples that can
be placed in memory
- Length (number of bytes) of each sample
- Number of casts in memory if in
Profiling mode; number of headers in
memory if in Moored mode
Example: Send GetSD to 19plus V2 (user input in bold, command used to modify parameter in parentheses)
getsd
<StatusData DeviceType = 'SBE19plus' SerialNumber = '01906003'>
<DateTime>2011-06-05T10:53:03</DateTime> [DateTime=]
<LoggingState>not logging</LoggingState>
<EventSummary numEvents = '0'/> [can clear with ResetEC]
<Power>
<vMain>10.1</vMain>
<vLith>8.9</vLith>
<iMain>61.9</iMain>
<iPump>20.6</iPump>
<iExt01>67.2</iExt01>
</Power>
<MemorySummary>
<Bytes>150</Bytes> [can clear with InitLogging]
<Samples>10</Samples> [can clear with InitLogging]
<SamplesFree>4386532</SamplesFree> [can clear with InitLogging]
<SampleLength>15</SampleLength>
<Profiles>1</Profiles> [can clear with InitLogging]
</MemorySummary>
</StatusData>
Notes:
The DS response contains similar
information as the combined
responses from GetSD and GetCD,
but in a different format.
Sending GetSD causes the pump to
turn on for a moment, so that the
19plus V2 can measure and output
the pump current. Because the
pump is designed to be water
lubricated, you will hear a noise
when the impeller spins in air.
Running the pump dry for such a
short time will not harm the pump.
In the example below, only voltage
channel 0 is enabled, so external
voltage current iext2345= (for
channels 2, 3, 4, and 5) is not
shown.
In the example below, no RS-232
sensor is enabled, so RS-232
sensor current iserial= is not shown.
If the 19plus V2 is set up with a WET
Labs ECO-FL fluorometer with
Bio-Wiper (Biowiper=Y) and if
OutputExecutedTag=Y, the GetSD
response shows: <Executing/>
to allow time for the Bio-Wiper to
close before it measures the
enabled external voltage currents.