1/14/2012
14
Recommended Default SBE 9 plus
D
ata Processing Parameters
• DATCNV (Module 1)
– Output up and downcasts of all parameters of interest. Only process on independent parameters
(T,C,P, OXVOLTS, Modulo Errors etc.)
– Output converted variables (salinity, DO concentration) if comparing to water samples
• ALIGNCTD (Module 3)
– SBE 11 usually advances C +0.073 secs
– Align DO and other sensor data to P and T accordingly in post processing
• FILTER (Module 2) only if continuous time series and no P outliers
– Pressure only 0.15 secs
• LOOPEDIT (Module 5)
– Only if ship heave a problem
– Select minimum fall speed according to data
• CellThermalMass (Module 4)
– ALWAYS
– Alpha = 0.03 and Tau = 7 secs
• DERIVE (Module 6)
– This is where you compute final Salinity, DO concentration, anything that is dependent on raw
measured and now processed variables (like T, C, P)
• Bin Average (Module 7)
– Do this AFTER running DERIVE