1/14/2012
14
Recommended Default SBE 9 plus
D
ata Processing Parameters
DATCNV (Module 1)
Output up and downcasts of all parameters of interest. Only process on independent parameters
(T,C,P, OXVOLTS, Modulo Errors etc.)
Output converted variables (salinity, DO concentration) if comparing to water samples
ALIGNCTD (Module 3)
SBE 11 usually advances C +0.073 secs
Align DO and other sensor data to P and T accordingly in post processing
FILTER (Module 2) only if continuous time series and no P outliers
Pressure only 0.15 secs
LOOPEDIT (Module 5)
Only if ship heave a problem
Select minimum fall speed according to data
CellThermalMass (Module 4)
ALWAYS
Alpha = 0.03 and Tau = 7 secs
DERIVE (Module 6)
This is where you compute final Salinity, DO concentration, anything that is dependent on raw
measured and now processed variables (like T, C, P)
Bin Average (Module 7)
Do this AFTER running DERIVE