24
How to Measure Pressure
Of
fset in the Lab Using a
Barometer
With offset in .con or .xmlcon file set to 0.0, pressure
measured by CTD should equal barometric pressure
Calculate offset (db) =
barometer reading – CTD reading
Conversion of psia to decibars:
decibars = (psia - 14.7) * 0.6894759
Enter calculated offset in .con or .xmlcon file
Example:
CTD reads -2.5 dbars
Ba
rometer reads 14.65 psia.
Converting to decibars, barometer reads
(14.65 - 14.7) * 0.6894759 = -0.034 dbars
offset (db) = barometer reading – CTD reading
= -0.034 - (-2.5) = 2.466