Section 4: Setting Up MicroTSG SBE 45
29
Operating Commands
Operating commands configure the MicroTSG’s response upon waking up,
and direct the MicroTSG to sample data once or at pre-programmed intervals.
Interval=x x= interval (seconds) between samples
(maximum 32767 seconds). MicroTSG
samples at this interval, and does not enter
quiescent (sleep) state between samples.
Minimum time between samples determined by
NCycles=, desired calculated parameters
(salinity, etc.), and baud rate; see Sample
Timing and Baud Rate, Cable Length, Power,
and Data Transmission Rate
in Section 2:
Description of MicroTSG.
AutoOff=x (Functional only if J1 jumper on PCB is in
Normal position)
x=Y: Go to sleep (enter quiescent state) if
2 minutes have elapsed without receiving a
command or without sampling data.
x= N: Do not automatically go to sleep.
AutoRun=x x=Y or N: Interacts with SingleSample= and
J1 jumper setting, as described in
table below.
SingleSample=x x=Y or N: Interacts with AutoRun= and J1
jumper setting, as described in table below.
Go Start sampling, as defined by SingleSample=
and Interval=.
Only applicable if:
AutoRun=N, or
AutoRun=Y and you previously sent Stop
to stop sampling.
Stop Stop sampling. Press Enter key to get
S> prompt before entering Stop.
J1 Jumper AutoRun SingleSample Effect
N
Y or N
Wake up (when Connect on Toolbar clicked or Enter key pressed while asleep) state
and wait for command.
Y N
Wake up (when Connect on Toolbar clicked or Enter key pressed while asleep) and
sample at rate specified by Interval=.
To stop sampling and get S> prompt, type Stop and press Enter key.
Normal
(pins 2
and 3)
Y Y
Wake up (when Connect on Toolbar clicked or Enter key pressed while asleep), take
and output single sample, and automatically go to sleep. To wake up and get S>
prompt, type Stop and press Enter key. Referred to as Serial Line Sync Mode.
N
Y or N Wake up when power applied and wait for command.
Y N
Wake up when power applied and sample at rate specified by Interval= until power
removed. Required settings for MicroTSG in 3-wire (power, ground, and
transmit) configuration or when using MicroTSG with Interface Box.
Autopower
(pins 1
and 2)
Y Y
Wake up when power applied and take and output a single sample. Wait for another
command until power removed.
Note:
If the MicroTSG is sampling data and
the external voltage is less than
6.15 volts for ten consecutive scans,
the MicroTSG halts logging.
Note:
If the total time required for the sample
is greater than Interval=, the
MicroTSG begins the next sample as
soon as it finishes transmitting the
current sample.
Note:
You may need to send Stop several
times to get the MicroTSG to respond.
This is most likely to occur if sampling
with a small Interval=.